[A new device for measuring occlusion pressure]

J Belda1, J Giner, P Casán

  • 1Departamento de Neumología, Hospital de la Santa Creu i Sant Pau., Barcelona.

Archivos De Bronconeumologia
|October 1, 1996
PubMed
Summary
This summary is machine-generated.

Related Concept Videos

You might also read

Related Articles

Articles linked to this work by shared authors, journal, and citation graph.

Sort by
Same author

Emerging Hyalomma lusitanicum: From identification to vectorial role and integrated control.

Medical and veterinary entomology·2023
Same author

Traumatic brain injury in the new millennium: new population and new management.

Neurologia·2022
Same author

Impact of Diabetes Mellitus and Frailty on Long-Term Outcomes in Elderly Patients with Acute Coronary Syndromes.

The journal of nutrition, health & aging·2020
Same author

First report on natural infection with Leishmania infantum in a domestic ferret (Mustela putorius furo) in Spain.

Veterinary parasitology, regional studies and reports·2020
Same author

Validation study of the dynamic parameters of pulse wave in pulmonary resection surgery.

Revista espanola de anestesiologia y reanimacion·2020
Same author

Assessing Adherence by Combining the Test of Adherence to Inhalers With Pharmacy Refill Records.

Journal of investigational allergology & clinical immunology·2019
Same journal

Advances in Sarcoidosis.

Archivos de bronconeumologia·2026
Same journal

Spanish Clinical Practice Guidelines for the Diagnosis and Management of Alpha-1 Antitrypsin Deficiency: 2026 Update.

Archivos de bronconeumologia·2026
Same journal

Thoracic Sympathectomy for Harlequin Syndrome.

Archivos de bronconeumologia·2026
Same journal

Prognostic Significance of Electrocardiographic Findings in Hemodynamically Stable Patients With Acute Pulmonary Embolism: Insights From the RIETE Registry.

Archivos de bronconeumologia·2026
Same journal

Endothelial and Angiogenic Biomarkers in Obstructive Sleep Apnea: Longitudinal Associations With Continuous Positive Airway Pressure.

Archivos de bronconeumologia·2026
Same journal

Robotic Resection of Ectopic Paratracheal Thymoma: A Case Report.

Archivos de bronconeumologia·2026
See all related articles

A new automated device accurately measures mouth occlusion pressure (P0.1), a key indicator of respiratory muscle drive. This validated tool enhances the clinical application of P0.1 measurements in lung function analysis.

Area of Science:

  • Respiratory Physiology
  • Medical Device Validation

Background:

  • Mouth occlusion pressure (P0.1) is a valuable indirect measure of central respiratory drive.
  • Limited availability of validated automated devices hinders widespread P0.1 measurement.
  • Existing laboratory methods are often custom-built and not commercially available.

Purpose of the Study:

  • To validate a new automated manometer for measuring P0.1.
  • To assess the precision and accuracy of the novel device.
  • To compare the new device against a standard laboratory apparatus.

Main Methods:

  • A new semiconductor-based pressure transducer device was developed.
  • Linearity, accuracy, and sensitivity were tested using water columns (5 and 25 cmH2O).
  • Clinical validation involved 33 patients undergoing lung function tests, with five P0.1 measurements per patient.

Related Experiment Videos

Main Results:

  • The new device demonstrated high sensitivity (0.1 cmH2O) and linearity (r = 0.999).
  • Precision was excellent, with a coefficient of variation of 0%.
  • The new device showed a small, clinically insignificant bias compared to the standard apparatus (mean difference 0.11 cmH2O).

Conclusions:

  • The new automated manometer provides valid and precise P0.1 measurements.
  • The device's accuracy and ease of use support its clinical adoption.
  • Increased availability of such devices can promote wider use of occlusion pressure measurements in respiratory diagnostics.