Expired CO2 Measurement in Intubated or Spontaneously Breathing Patients from the Emergency Department
Published on: January 29, 2011
Continuous end-tidal carbon dioxide monitoring during normofrequent jet-ventilation
L F Bach1, H Wanner-Olsen, B N Andersen
1Department of Anaesthesiology, Viborg Hospital, Denmark.
Background:
Continuous end-tidal carbon dioxide (ETCO2) monitoring during normofrequent jet-ventilation (NFJV) has not previously been successful and no correspondence between ETCO2 and arterial carbon dioxide (PaCO2) demonstrated.
Methods:
During NFJV jet-ventilation in 19 healthy volunteers, continuous ETCO2 was measured and compared to PaCO2 values. The original ETCO2 sampling line from our Datex CO2 measuring equipment was placed longitudinally against a suction catheter and both were wrapped with aluminium foil. They were placed 4-5 cm above the carina and the position was verified with a fiberoptic bronchoscope.
Results:
The correlation between methods gave a Pearson's product moment correlation (r value) of 0.836, significant at the P < 0.001 level, indicating a close correlation between methods. A difference against mean scatter diagram confirmed that ETCO2 and PaCO2 measurements are of equal value in monitoring healthy patients during NFJV.
Conclusion:
A valuable method of continuous ETCO2 monitoring during NFJV is presented.
Related Concept Videos
Assessment of Ventilation I: Respiratory Rate
A Ventilation assessment is critical for monitoring a patient's health status. Respiration, one of the most accessible vital signs, provides insights into the function of numerous body systems and can indicate serious health issues, such as brainstem injuries from head trauma.
Critical Guidelines for Assessing Ventilation:
Assessment of Diffusion and Perfusion
The Role of Diffusion in Respiration
Diffusion is the process by which molecules move from an area of higher concentration to an area of lower concentration. In the respiratory system, this principle...
Assessment of Ventilation II: Respiratory Depth and Rhythm
Respiratory depth measures the volume of air inhaled or exhaled during a breath. It can vary from shallow to deep and typically remains consistent when a person is at rest or asleep. Occasionally, individuals will automatically inhale deeply, known as sighing, which inflates the lungs with more air than normal breathing.
To assess respiratory depth, observe the degree of chest excursion or movement:
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Mechanical Ventilation III: Noninvasive Ventilation
Noninvasive Positive-Pressure Ventilation (NIPPV)
Ventilatory Modes
There are three ventilatory modes: full support, partial support, and spontaneous. These are described below.
Full Support Modes
Full support modes include controlled mechanical ventilation, continuous mandatory...