Assessing contrast on MR images

S D Wolff1, R S Balaban

  • 1Laboratory of Cardiac Energetics, National Heart, Lung, and Blood Institute, National Institutes of Health, Bethesda, MD 20892-1061, USA.

Radiology
|January 1, 1997
PubMed
Summary
This summary is machine-generated.

Related Concept Videos

You might also read

Related Articles

Articles linked to this work by shared authors, journal, and citation graph.

Sort by
Same author

Parallel transmit excitation at 1.5 T based on the minimization of a driving function for device heating.

Medical physics·2015
Same author

Use of independent component analysis to improve signal-to-noise ratio in multi-probe fluorescence microscopy.

Journal of microscopy·2014
Same author

Motion compensation for in vivo subcellular optical microscopy.

Journal of microscopy·2014
Same author

Interpretation of ³¹P NMR saturation transfer experiments: what you can't see might confuse you. Focus on "Standard magnetic resonance-based measurements of the Pi→ATP rate do not index the rate of oxidative phosphorylation in cardiac and skeletal muscles".

American journal of physiology. Cell physiology·2011
Same author

Rapid overlapping-volume acquisition and reconstruction (ROVAR): automated 3D tiling for high-resolution, large field-of-view optical microscopy.

Journal of microscopy·2011
Same author

Optimizing multiphoton fluorescence microscopy light collection from living tissue by noncontact total emission detection (epiTED).

Journal of microscopy·2010
Same journal

Erratum for: Prediction of Lobar Emphysema Progression with a CT-Based Foundational Model.

Radiology·2026
Same journal

Erratum for: Associations of MRI-derived Paraspinal IMAT and LMM with Cardiometabolic Risk Factors: Results from a German Cohort.

Radiology·2026
Same journal

Erratum for: Blue Rubber Bleb Nevus Syndrome.

Radiology·2026
Same journal

Redefining the Clinical Role of MRI in Endometrial Cancer Staging.

Radiology·2026
Same journal

To Ablate or Not to Ablate: The Colorectal Liver Metastasis Question.

Radiology·2026
Same journal

The Limits of Radiologic Categorization in Pulmonary Nonsolid Nodules.

Radiology·2026
See all related articles

Signal difference-to-noise ratio (SDNR) is a display-independent metric for assessing magnetic resonance imaging (MRI) pulse sequences. SDNR should be preferred over contrast-to-noise ratio (CNR) for objective analysis of tissue discrimination.

Area of Science:

  • Medical Imaging
  • Radiology
  • Biophysics

Background:

  • Magnetic resonance imaging (MRI) pulse sequences are crucial for differentiating between pathological and normal tissues.
  • Objective analysis of MRI images typically relies on signal intensity measurements, but standardized methods are lacking.
  • Existing literature shows considerable disagreement on the optimal approach for performing signal intensity analysis in MRI.

Purpose of the Study:

  • To evaluate the suitability of contrast-to-noise ratio (CNR) and signal difference-to-noise ratio (SDNR) for assessing MRI pulse sequences.
  • To determine the preferred objective parameter for evaluating the contrast-generating ability of MRI pulse sequences.
  • To discuss the value and limitations of CNR and SDNR in MRI image analysis.

Main Methods:

Related Experiment Videos

  • Discussion of the definition and calculation of contrast-to-noise ratio (CNR).
  • Explanation of the display-dependent nature of CNR, requiring measurement of photon intensity reaching the observer.
  • Introduction of signal difference-to-noise ratio (SDNR) as a display-independent parameter reflecting pulse sequence contrast generation.

Main Results:

  • Contrast-to-noise ratio (CNR) is display-dependent, limiting its use for objective sequence comparison.
  • Signal difference-to-noise ratio (SDNR) is a display-independent metric that accurately reflects the intrinsic contrast-generating capability of an MRI pulse sequence.
  • SDNR is proportional to CNR under specific display conditions, making it a reliable surrogate for comparison.

Conclusions:

  • Signal difference-to-noise ratio (SDNR) is the preferred parameter for objectively assessing and comparing the contrast-generating performance of MRI pulse sequences.
  • SDNR's display-independent nature ensures consistent and reproducible evaluation of MRI techniques.
  • Understanding the limitations of both CNR and SDNR is essential for accurate interpretation of MRI image quality.