Quantitative detection of outlet strut separations in Björk-Shiley convexo-concave mitral valves

K A Powell1, N A Obuchowski, K D Mueller

  • 1Department of Biomedical Engineering, Cleveland Clinic Foundation, OH 44195, USA. powell@bme.ri.ccf.org

Circulation
|December 15, 1996
PubMed

Insights

A new objective postprocessing technique evaluates Björk-Shiley convexo-concave heart valve strut integrity. This method aids in identifying valves at risk of fracture, improving patient safety and diagnostic accuracy for outlet strut fractures.

Area of Science:

  • Biomedical Engineering
  • Cardiovascular Imaging
  • Materials Science

Background:

  • Björk-Shiley convexo-concave (BSCC) heart valves have a history of catastrophic outlet strut fractures (OSFs).
  • Approximately two-thirds of reported OSFs resulted in patient mortality.
  • Single-leg separation (SLS) precedes complete OSF, indicating a critical failure point.

Purpose of the Study:

  • To develop an objective postprocessing technique for evaluating BSCC heart valve outlet strut integrity.
  • To quantitatively assess strut leg integrity using cineradiographic images.
  • To identify BSCC heart valves at risk of complete OSF.

Main Methods:

  • Acquired cineangiographic images from 22 individuals (12 intact, 10 SLS valves).
  • Verified valve status ex vivo.
  • Evaluated quantitative measures of SLS, including decrease in pixel intensity (DIPI) ratio and gap half-width.

Main Results:

  • Developed an objective postprocessing technique for evaluating outlet strut integrity.
  • DIPI ratio and gap half-width were identified as diagnostic metrics for SLS.
  • The model based on DIPI ratio achieved an area under the receiver operating characteristic curve of 0.892.

Conclusions:

  • An objective postprocessing technique for evaluating BSCC heart valve strut integrity in cineradiographic images has been developed.
  • The technique provides objective and quantitative assessment of strut integrity.
  • The objective review demonstrated sensitivity comparable to subjective expert review at high specificity.
Abstract

Related Concept Videos

Mitral Valve Prolapse I: Introduction01:27

Mitral Valve Prolapse I: Introduction

IntroductionThe mitral valve, one of the heart's four valves, regulates blood flow. These valves have flaps that open and close to direct blood properly through the heart and body. During each heartbeat, the flaps open for blood to pass through and seal shut to prevent backflow. Specifically, the mitral valve opens to allow blood flow from the heart's upper left chamber to the lower left chamber. It then closes securely as the lower left chamber contracts to pump blood to the body, preventing...
Mitral Valve Prolapse II: Assessment and Management01:22

Mitral Valve Prolapse II: Assessment and Management

IntroductionA range of clinical features characterizes Mitral Valve Prolapse (MVP), but it is important to note that many individuals with MVP are asymptomatic and may remain so throughout their lives. For those who do exhibit symptoms, the following are the key clinical features:Palpitations: This is a common symptom where individuals feel an irregular or rapid heartbeat. Palpitations in MVP are often due to arrhythmias such as premature ventricular contractions or supraventricular tachycardia.
Mitral Stenosis II: Clinical features and Diagnostic Tests01:23

Mitral Stenosis II: Clinical features and Diagnostic Tests

Mitral stenosis is a heart condition in which the mitral valve, which allows blood to flow from the left atrium to the left ventricle, becomes narrowed or stenotic. This narrowing hinders blood flow and leads to clinical symptoms requiring specific medical evaluations and management strategies. The following overview outlines the clinical symptoms, assessments, diagnostic findings, prevention methods, and treatments for mitral stenosis.Clinical ManifestationsDyspnea (shortness of breath): This...