EIT data noise evaluation in the clinical environment
Related Experiment Video
Updated: Oct 9, 2026

Troubleshooting and Quality Assurance in Hyperpolarized Xenon Magnetic Resonance Imaging: Tools for High-Quality Image Acquisition
Published on: January 5, 2024
S Meeson1, B H Blott, A L Killingback
1Department of Physics, University of Southampton, UK.
Abstract:
In the clinical environment the reliable interpretation of EIT images depends on the quality of the data. In the electrically noisy hospital environment the system performance needs to be assessed for each clinical investigation. From the model of noise presented, a figure of merit for comparisons of system performance with a known standard, or with previous studies, can be generated. The method depends on calculating the variances of the differences in reciprocity measurements as a function of the distance between the current drive electrodes and the receive voltage electrodes. These measurements fit the noise model, with minimal interference from physiological variability, and permit a figure of merit to be calculated which is a representation of the noise at the point to the system. Typical figures of merit are 7.36 +/- 0.03 microV for a test card and 10.50 +/- 16 microV for subject data.