Related Experiment Video
Updated: Aug 10, 2026

Quantification of the Immunosuppressant Tacrolimus on Dried Blood Spots Using LC-MS/MS
Published on: November 8, 2015
Direct assay of tinidazole in human serum by micellar liquid chromatography
D Habel1, S Guermouche, M H Guermouche
1Insitut de Chimie, USTHB, Alger, Algeria.
Abstract:
A liquid chromatographic procedure for the direct determination of tinidazole in human serum is presented. It includes the use of a micellar mobile phase consisting of SDS (5.10(-2) M): propan-1-ol; (94:6, v/v) and a mu Bondapak CN column with UV detection at 320 nm. No solvent extraction or deproteinization are necessary. The linearity (0.1-10 mg L), the precision (3%), the reproducibility (1.3%), the recovery (99%), and the detection limit (0.1 mg L) in the tinidazole determination are comparable and sometimes greater than the corresponding tinidazole parameters when deproteinization and conventional reversed-phase HPLC are used. One hundred injections of serum samples do not affect the column life. The procedure is applied to ascertain the pharmacokinetics of 10 mg/kg of tinidazole.