Duplex ultrasound criteria for the identification of carotid stenosis should be laboratory specific
K M Kuntz1, J F Polak, A D Whittemore
1Department of Medicine, Brigham and Women's Hospital, Boston, Mass., USA.
Background And Purpose:
Published criteria for the determination of carotid stenosis have been widely applied by vascular laboratories. We compared two vascular laboratories and their duplex ultrasound (DU) machines in terms of their overall diagnostic performance and the optimal criteria to identify patients who have a 70% to 99% stenosis of the internal carotid artery.
Methods:
Measurements of stenosis by DU and angiography were compared for 123 carotid arteries (60 arteries, laboratory A; 63 arteries, laboratory B). Receiver operating characteristic (ROC) curves were created, and the areas under the ROC curves and the optimal criteria for determining a 70% to 99% stenosis were compared. Multiple regression analysis was used to measure the effect of laboratory on the relationship between angiographic stenosis and DU velocity parameters.
Results:
Areas under the ROC curves were similar for both laboratories (0.89 to 0.90, laboratory A; 0.90 to 0.92, laboratory B). However, the optimal criterion for the identification of a 70% to 99% carotid stenosis was different for each laboratory. For most velocity parameters, based on regression analyses, the predicted percent angiographic stenosis for laboratory A was significantly greater than that for laboratory B. In addition, performance differed between the laboratories when established criteria from the literature were applied.
Conclusions:
Two vascular laboratories with similar diagnostic accuracy by ROC analysis have markedly different "optimal" DU criteria. For a given angiographic stenosis, velocities in one laboratory were consistently greater than those in the other laboratory. Laboratory-specific criteria rather than published criteria should be used to identify patients with internal carotid artery stenoses.
More Related Videos
09:33A Methodological Approach to Non-invasive Assessments of Vascular Function and Morphology
Published on: February 7, 2015
09:36A Magnetic Resonance Imaging-based Computational Protocol for Analysis of Plaque Morphology and Hemodynamics in Patients with Carotid Artery Stenosis
Published on: August 12, 2025
Related Concept Videos
Atherosclerosis II: Clinical Manifestations and Diagnostic Tests
Peripheral Arterial Disease II: Clinical Manifestations and Diagnostic Evaluation