Related Experiment Video
Updated: Sep 21, 2026

Use of a Low-flow Digital Anesthesia System for Mice and Rats
Published on: September 7, 2016
Ether in an isoflurane vaporizer and the use of vapour analysers in safe anaesthesia
1Department of Anaesthesia, Royal Hallamshire Hospital, Sheffield.
Abstract:
A Penlon Sigma isoflurane vaporizer was found to contain some diethyl ether. This had escaped from a ruptured temperature compensation device within the vaporizer. The Datex AS/3 vapour analyser detected the foreign agent and displayed a warning. Different types of vaporizer compensation devices are highlighted and the use of vapour analysers is discussed. We conclude that these analysers are invaluable for indicating that the patient is receiving the correct concentration of an inhalation agent but are of little use in the detection of all but a few possible contaminants.