Objective:
This study evaluated the accuracy of placing right atrial catheters using an electrocardiographic (ECG) technique.
Summary Background Data:
Placement of right atrial catheters for vascular access is a common operative procedure. Accurate placement is essential for proper function. Previous placement techniques have used fluoroscopy, which is both time consuming and hazardous.
Methods:
The accuracy of placement of 1236 right atrial catheters using an ECG technique was compared to placement of 586 catheters using fluoroscopy between March 1991 and November 1995. In the ECG technique, the catheter was flushed with sodium bicarbonate. A sterile left-leg ECG lead was attached to the catheter with the other ECG leads applied normally. On advancing the catheter through the superior vena cava, the P-wave amplitude (lead II) increased in negative deflection until greater than the QRS complex. Passing the sinoatrial node, the P-wave developed an initial positive then negative deflection. The catheter was positioned so the P-wave was biphasic, representing a position midway between the sinoatrial and atrioventricular nodes. For the fluoroscopic technique, catheters were positioned under direct observation just within the atrium estimated from cardiac contour. Use of contrast was optional if atrial anatomy was unclear.
Results:
Postoperative portable chest x-rays showed the ECG method to position the catheter tip within the right atrium just as accurately (average, 1.9 +/- 1.3 cm) as with the use of fluoroscopy (average, 1.1 +/- 1.6 cm). The ECG method eliminated an average of 20 seconds of radiation exposure, an average of 3.0 minutes operating room time (p < 0.04), avoided all risks of contrast dye, and saved $279.10 per case.
Conclusions:
The ECG method is a satisfactory alternative to that of fluoroscopy for placement of long-term central venous catheters into the right atrium.
Related Concept Videos
Electrocardiogram
Three major waveforms are present in a typical ECG recording: the P wave, the QRS complex, and the T...
Electrocardiogram Fundamentals
An electrocardiogram (ECG) is a diagnostic tool for identifying cardiac conditions such as arrhythmias, conduction abnormalities, and myocardial ischemia.
Definition
An electrocardiogram (ECG) visualizes the heart's electrical activity by tracing the electrical movement associated with each heartbeat on a graph or monitor. As the heart beats, an electrical wave passes through it, correlating with the cardiac cycle events.
Parts of an ECG
An ECG utilizes electrodes on the skin to...
Cardiac Catheterization I: Pre-Procedure Overview
Cardiac Catheterization II: Right Heart Catheterization
Cardiac Catheterization III: Left Heart Catheterization
Cardiopulmonary Resuscitation III: AED Use