Inaccuracies of nitric oxide delivery systems during adult mechanical ventilation
Background:
Various systems to administer inhaled nitric oxide (NO) have been used in patients and experimental animals. We used a lung model to evaluate five NO delivery systems during mechanical ventilation with various ventilatory patterns.
Methods:
An adult mechanical ventilator was attached to a test lung configured to separate inspired and expired gases. Four injection systems were evaluated with NO injected either into the inspiratory circuit 90 cm proximal to the Y piece or directly at the Y piece and delivered either continuously or only during the inspiratory phase. Alternatively, NO was mixed with air using a blender and delivered to the high-pressure air inlet of the ventilator. Nitric oxide concentration was measured from the inspiratory limb of the ventilator circuit and the tracheal level using rapid- and slow-response chemiluminescence analyzers. The ventilator was set for constant-flow volume control ventilation, pressure control ventilation, pressure support ventilation, or synchronized intermittent mandatory ventilation. Tidal volumes of 0.5 l and 1 l were evaluated with inspiratory times of 1 s and 2 s.
Results:
The system that premixed NO proximal to the ventilator was the only one that maintained constant NO delivery regardless of ventilatory pattern. The other systems delivered variable NO concentration during pressure control ventilation and spontaneous breathing modes. Systems that injected a continuous flow of NO delivered peak NO concentrations greater than the calculated dose. These variations were not apparent when a slow-response chemiluminescence analyzer was used.
Conclusions:
NO delivery systems that inject NO at a constant rate, either continuously or during inspiration only, into the inspiratory limb of the ventilator circuit produce highly variable and unpredictable NO delivery when inspiratory flow is not constant. Such systems may deliver a very high NO concentration to the lungs, which is not accurately reflected by measurements performed with slow-response analyzers.
More Related Videos
05:46A Novel Inhalation Mask System to Deliver High Concentrations of Nitric Oxide Gas in Spontaneously Breathing Subjects
Published on: May 4, 2021
08:23Chemiluminescence-based Assays for Detection of Nitric Oxide and its Derivatives from Autoxidation and Nitrosated Compounds
Published on: February 16, 2022
Related Concept Videos
Oxygen Delivering System I: Nasal Cannula and Face Mask
Nasal Cannula
A nasal cannula is a lightweight tube split at one end into two prongs and placed in the nostrils. It is typically used to deliver low to medium levels of oxygen.
Suggested flow rate: The suggested flow rate for a nasal cannula typically ranges between 1 and 6 L/min.
Oxygen percentage setting:...
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Administering Oxygen by Nasal Cannula
Nasal Cannulas
A nasal cannula is a lightweight tube split into two prongs placed in the nostrils, connected...
Mechanical Ventilation I: Indication and Settings
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...
Mechanical Ventilation III: Noninvasive Ventilation
Noninvasive Positive-Pressure Ventilation (NIPPV)