An improved capillary blood-filter paper-graphite furnace atomic absorption spectrometric method for lead screening
1Detroit Medical Center University Laboratories, Michigan 48201, USA.
Journal of Analytical Toxicology
|March 1, 1997
Summary
A revised filter paper method for analyzing capillary blood samples using graphite furnace atomic absorption spectrometry (GFAAS) shows high accuracy for lead screening. This easier procedure is suitable for widespread use, with minimal false positives or negatives.
Area of Science:
- Analytical Chemistry
- Environmental Health
- Clinical Diagnostics
Background:
- Accurate blood lead level (BLL) monitoring is crucial for public health, especially in children.
- Traditional venous blood sampling can be invasive and challenging for routine screening.
- Filter paper (FP) collection offers a less invasive alternative for blood sample collection.
Purpose of the Study:
- To describe and validate a revised procedure for processing capillary blood samples on filter paper (FP) for lead analysis using graphite furnace atomic absorption spectrometry (GFAAS).
- To assess the accuracy, reliability, and practicality of the FP method compared to traditional venous blood sampling for lead screening.
Main Methods:
- A revised protocol for processing dried capillary blood spots from filter paper (FP) using GFAAS was developed.
- A 1/4-inch diameter disk was punched from FP samples for analysis.
- Results from FP samples were compared with corresponding venous blood samples, analyzing variables like hematocrit and filter paper blanks.
Main Results:
- A high correlation (0.975) was observed between FP and venous blood lead results.
- The revised FP method demonstrated low false-positive (0.8%) and false-negative (1.7%) rates.
- No significant differences were found for samples with hematocrit values of 30-39%, and filter paper blanks were minimal (< 2 µg/dL without correction, 0.1 µg/dL with correction).
- Environmental lead contamination was not detected during a year-long urban sampling study.
Conclusions:
- The revised FP method is a simpler, accurate, and reliable technique for lead screening using capillary blood.
- Fingerstick capillary blood collection on filter paper is suitable for widespread lead screening programs.
- Adherence to proper contamination precautions is essential for accurate results.