Related Experiment Videos

Comparison of in vivo red cell survival of donations collected by Haemonetics MCS versus conventional collection

F Regan1, P Teesdale, S Garner

  • 1North London Blood Centre, Colindale, UK.

The Haemonetics Multicomponents System (MCS) cell separator allows concurrent donation of red cells in addition to platelets and/or plasma, thus increasing the versatility of apheresis donations. In vivo survival of autologous red cells obtained by MCS was compared with red cells collected conventionally. In this cross-over controlled study, five male volunteers donated one unit of red cells by MCS and one unit of whole blood by the conventional manual method, 3 months apart. After storing donations in SAG-M for 35 days under standard conditions, radioactive (51Cr)-labelled autologous red cells were injected into each donor. The post-transfusion recovery (PTR) of red cells at 24 and 48 h did not show any significant difference between red cells obtained manually and by MCS, indicating that processing differences have no detrimental effects on red cell survival.