Related Experiment Videos
Small-bowel obstruction and perforation. A rare complication of an esophageal stent
T H Henne1, B Schaeff, V Paolucci
1Department of Surgery, University of Frankfurt am Main, Theodor Stern Kai 7, D-60590 Frankfurt, Germany.
Surgical Endoscopy
|April 1, 1997
Abstract:
To bridge a malignant stenosis after esophageal cancer recurrence two silicon-coated wall stents were inplanted in a 52-year-old patient within 6 months. Two weeks after the second stent was implanted, clinical examination showed dislocation of the prosthesis. Intraoperatively the two stents were found sticking in the side-to-side jejunostomy of a former Billroth II operation, leading to perforation there.