Scanning laser Doppler flowmetry in glaucoma
G Holló1, T J van den Berg, E L Greve
1Glaucoma Center, University of Amsterdam, The Netherlands.
International Ophthalmology
|January 1, 1996
Summary
Heidelberg Retina Flowmeter measurements show no significant difference in capillary perfusion between healthy eyes and those with primary open angle glaucoma (POAG) or normal pressure glaucoma (NPG). Test-frame position is crucial for accurate retinal blood flow assessment.
Area of Science:
- Ophthalmology
- Medical Imaging
- Physiology
Background:
- Glaucoma is a leading cause of irreversible blindness.
- Assessing ocular blood flow is crucial for understanding glaucoma pathophysiology.
- Scanning laser Doppler flowmetry (SLDF) offers a non-invasive method to measure capillary perfusion.
Purpose of the Study:
- To evaluate the reproducibility and diagnostic potential of the Heidelberg Retina Flowmeter (HRF) for measuring retinal and optic disc capillary perfusion.
- To compare ocular blood flow in healthy volunteers, primary open angle glaucoma (POAG) patients, and normal pressure glaucoma (NPG) patients.
Main Methods:
- The Heidelberg Retina Flowmeter was used to measure retinal and neuroretinal rim capillary perfusion in 31 healthy volunteers, 42 POAG patients, and 17 NPG patients.
- Measurements were taken on the temporal retina, nasal retina, and optic disc rim.
- Reproducibility and correlation analyses were performed.
Main Results:
- Optic disc flow was significantly higher than retinal flow (p < 0.000001).
- Reproducibility varied by location and group, with NPG showing good reproducibility (10-13%).
- No statistically significant differences in perfusion were found between healthy controls and glaucoma patients (POAG or NPG).
- Intraocular pressure and treatment type did not influence retinal flow.
- Significant variation in flow was observed between adjacent frame positions (18.5%), highlighting the importance of test-frame positioning.
Conclusions:
- The Heidelberg Retina Flowmeter demonstrated variable reproducibility, particularly in healthy controls.
- No significant differences in capillary perfusion were detected between healthy eyes and glaucomatous eyes using this method.
- Accurate positioning of the test-frame is critical for reliable measurements of retinal blood flow.