Programming of implantable cardioverter-defibrillators on the basis of the upper limit of vulnerability
Related Experiment Video
Updated: Aug 1, 2026

A New Single Chamber Implantable Defibrillator with Atrial Sensing: A Practical Demonstration of Sensing and Ease of Implantation
Published on: February 28, 2012
C D Swerdlow1, C T Peter, R M Kass
1Division of Cardiology, Cedars-Sinai Medical Center, Los Angeles, Calif, USA. swerdlow@ucla.edu
Background:
A patient-specific measure of defibrillation efficacy that requires a minimum number of ventricular fibrillation (VF) episodes would be valuable for programming implantable cardioverter-defibrillators (ICDs). The upper limit of vulnerability (ULV) is the weakest shock strength at or above which VF is not induced when a stimulus is delivered during the vulnerable phase of the cardiac cycle. It correlates with the defibrillation threshold (DFT) and can be determined with a single episode of VF. The objective of this study was to test the hypothesis that ICDs programmed on the basis of the ULV convert spontaneous ICD-detected VF reliably.
Methods And Results:
We studied 100 consecutive patients at ICD implantation and during follow-up of 20 +/- 7 months. At implantation, the ULV and DFT were determined, and the ICD system was tested at a shock strength equal to the ULV + 3 J. During follow-up, the strength of the first shock was programmed to the ULV + 5 J for arrhythmias detected in the VF zone (cycle length < 292 +/- 17 ms). We reviewed stored detection intervals and electrograms from spontaneous episodes of ICD-detected VF to determine the success rate for appropriate first shocks. The programmed first-shock strength was 17.5 +/- 5.2 J. During follow-up, there were 120 appropriate first shocks in 37 patients. The arrhythmia was rapid monomorphic ventricular tachycardia (VT) in 70% of episodes (31 patients), VF in 11% (13 patients), polymorphic VT in 1%, and unclassified in 17% (15 patients). The first shock was successful in 119 of 120 episodes (99%; 95% CI, 93% to 100%). One unclassified episode required two shocks. No patient had syncope associated with an ICD shock or arrhythmic death.
Conclusions:
ICD shocks can be programmed on the basis of the ULV, a measurement made in regular rhythm, without a direct measure of defibrillation efficacy.
More Related Videos
06:57Ablation of Ischemic Ventricular Tachycardia Using a Multipolar Catheter and 3-dimensional Mapping System for High-density Electro-anatomical Reconstruction
Published on: January 31, 2019
08:43Advanced Cardiac Rhythm Management by Applying Optogenetic Multi-Site Photostimulation in Murine Hearts
Published on: August 26, 2021
Related Concept Videos
Disturbances in Heart Rhythm
Arrhythmias are categorized by their speed, rhythm, and origin. A slow heart...
Dysrhythmias VI: Management of Dysrhythmias
Cardiopulmonary Resuscitation III: AED Use
Cardiomyopathy V: Interprofessional Care
ECG Interpretation of Arrhythmias II: Atrial, Junctional and Ventricular Arrhythmias