Basophil counts on the technicon H*1 automated counter
Clinical and Laboratory Haematology
|March 1, 1996
Summary
The Technicon H*1 automated blood cell counter offers improved precision for basophil counts but requires microscopic confirmation for elevated results due to pseudo-basophilia. This automated method provides clinical usefulness and stability for basophil analysis.
Area of Science:
- Hematology
- Clinical Pathology
- Medical Laboratory Science
Background:
- Automated blood cell counters are essential for routine hematological analysis.
- Accurate basophil counting is clinically relevant but challenging with manual methods.
- The Technicon H*1 instrument utilizes specific technology for basophil enumeration.
Purpose of the Study:
- To evaluate the precision, accuracy, stability, and clinical utility of the Technicon H*1 basophil count.
- To establish a reference range for absolute basophil counts using this automated system.
- To assess the reliability of automated basophil counts in identifying true basophilia versus pseudo-basophilia.
Main Methods:
- Assessed precision and accuracy of the Technicon H*1 basophil count against established criteria.
- Determined the stability of basophil counts at room temperature and 4 degrees C over 3 days.
- Derived a reference range for absolute basophil counts.
- Compared automated counts with manual microscopic examination to identify discrepancies.
Main Results:
- Precision was superior to manual counts but lower than expected based solely on random error.
- Accuracy was sufficient for clinical use, though 'pseudo-basophilia' frequently caused elevated counts.
- Basophil counts remained stable for up to 3 days under tested conditions.
- Established a reference range of 0.02-0.09 x 10(9)/L.
- Elevated counts serve as an alert for abnormal samples.
Conclusions:
- The Technicon H*1 provides clinically useful basophil counts, but microscopic confirmation is vital for diagnostically significant elevated results.
- Pseudo-basophilia is a common factor in elevated automated basophil counts, necessitating careful interpretation.
- The findings are generalizable to other instruments in the Technicon H series (H*2, H*3) due to shared technology.