Comparison of gastric air tonometry with standard saline tonometry
Related Experiment Video
Updated: Nov 2, 2026

Ultrasonography-Guided Nasogastric Tube Placement and Verification: A Standardized Nursing Protocol
Published on: April 7, 2026
G Tzelepis1, V Kadas, A Michalopoulos
1Surgical Intensive Care Unit, Onassis Cardiac Surgical Center, Athens, Greece.
Objective:
To compare partial pressure of carbon dioxide (PCO2) measurements obtained by sampling gastric intraluminal air with those obtained by standard saline tonometry.
Design:
Prospective, unblinded study.
Setting:
Intensive care unit in a tertiary cardiac surgical center.
Patients:
20 patients undergoing cardiac surgery.
Interventions:
Gastric tonometric catheters were inserted, gastric fluid was aspirated, and 100 cc of air was injected into the stomach.
Measurements:
After an equilibration period of 30 min, samples of gastric air and saline were anaerobically aspirated and analyzed on a standard blood gas machine. The reproducibility of PCO2 measurements in a given patient was assessed by analyzing consecutive samples of gastric air and calculating the coefficient of variation (CV).
Results:
PCO2 values measured in samples of gastric air (PCO2 air) were highly correlated with those derived by saline tonometry (PCO2ss)(r2 = 0.95, p = 0.0001); PCO2 air was significantly greater than PCO2ss (50 +/- 17 vs 48 +/- 17 mmHg, p = 0.0001). Intramucosal pH (pHi) calculated from PCO2 air was significantly lower than that calculated from PCO2ss (7.26 +/- 0.23 vs 7.28 +/- 0.24, p = 0.0001). Analysis of intermethod differences showed significant bias for both PCO2 (2.4 +/- 7.6 mmHg, mean +/- 2SD, bias +/- precision) and pHi ( -0.023 +/- 0.074, mean +/- 2SD, bias +/- precision). The within-subject variability of replicate PCO2 measurements in gastric air was low (CV = 2.6 +/- 0.8).
Conclusion:
We conclude that intraluminal PCO2 can be accurately determined in postoperative cardiac surgery patients by instilling air into the stomach and analyzing samples of gastric air on a standard blood gas machine, In comparison with saline tonometry, air tonometry consistently yields lower pHi values.
Related Concept Videos
Enteral Nutrition I: Orogastric and Nasogastric Feeding
Orogastric (OG) and nasogastric (NG) feeding are two standard methods used for enteral nutrition. Enteral nutrition is often preferred over...
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...