Experimental correlation-based identification of X-ray CT point spread function. Part 1: Method and experimental results
S Doré1, R E Kearney, J A De Guise
1Ecole de Technologie Supérioure, Montréal, Québec, Canada. sdore@mec.etsmtl.ca
Abstract:
Knowledge of the point spread function (PSF) of an imaging system is important when studying the characteristics of the blur present in the images. Published experimental PSF identification techniques adapt classical one-dimensional linear system identification strategies using impulse, step function or periodic input signals. This study proposes and successfully applies a correlation method based on the Wiener-Hopf equation to identify the PSF of a CT scanner. The input consists of a series of pseudorandomly located holes. Results are found to be statistically equivalent to those obtained with the impulse method at a 90% two-sided confidence interval. Like the impulse method, it readily yields two-dimensional estimate, but the larger input circumvents the major objection to the use of a wire input. Furthermore, it is relatively immune to output noise, offering an advantage over edge methods. This resistance to noise may prove helpful for nuclear medicine imaging techniques, for which the signal-to-noise ratio is much lower than that X-ray for CT.
Related Concept Videos
X-ray Crystallography
Diffraction
Diffraction is the change in the direction of travel experienced by an electromagnetic wave when it encounters a physical barrier whose dimensions are comparable to those of the wavelength of the light. X-rays are electromagnetic radiation with wavelengths about as long as the distance between neighboring...
X-ray Diffraction of Biological Samples
According to Bragg's law, when X-rays strike the sample positioned on a stage, the rays are scattered by the electron clouds around the sample atoms. The X-ray diffraction or scattering is caused by constructive interference of the X-ray waves that reflect off the internal crystal...