Computer versus manual measurement of ST-segment deviation

M M Pelter1, M G Adams, B J Drew

  • 1University of California, San Francisco 94143, USA.

Journal of Electrocardiology
|April 1, 1997
PubMed
Summary

Computerized electrocardiogram (ECG) ST scores correlate highly with manual measurements, but computers detect subtle ischemia missed by human analysis. Manual measurements show high interrater reliability, though computer and manual scores should not be combined in research.

Related Concept Videos

You might also read

Related Articles

Articles linked to this work by shared authors, journal, and citation graph.

Sort by
Same author

Severe perineal laceration during operative vaginal delivery: the impact of occiput posterior position.

Journal of perinatology : official journal of the California Perinatal Association·2014
Same author

Family Planning: A Survey of Clinic Patients.

The Ulster medical journal·2010
Same author

Simulated consumer exposure to propellant HCFC 22 (chlorodifluoromethane) in aerosol personal products.

International journal of cosmetic science·2009
Same author

Progressive evaluation of skin irritancy of cosmetics using human volunteers.

International journal of cosmetic science·2009
Same author

Simulated consumer exposure to dimethyl ether and propane/butane in hairsprays.

International journal of cosmetic science·2009
Same author

Dose response to intrathecal diamorphine for elective caesarean section and compliance with a national audit standard.

International journal of obstetric anesthesia·2006

Area of Science:

  • Cardiology
  • Medical Technology
  • Biomedical Engineering

Background:

  • Total ST scores from electrocardiograms (ECGs) estimate ischemic burden and predict reperfusion.
  • Computerized systems offer higher precision (10 microV) than human analysis (50 microV) for ST deviation measurements.

Purpose of the Study:

  • Compare computer-derived versus manually derived ST scores in ischemic 12-lead ECGs.
  • Assess interrater reliability of manual ST score measurements between two experts.

Main Methods:

  • Analyzed 58 subjects with ≥100 microV ST deviation during percutaneous transluminal coronary angioplasty.
  • Measured ST deviation at J+80 ms across all 12 leads, comparing computer and blinded manual assessments.
  • Used paired t-tests and Pearson correlation coefficients for statistical analysis.

Related Experiment Videos

Main Results:

  • High correlation observed between computer and manual ST scores (r = .96).
  • Excellent interrater reliability between two manual reviewers (r = .96).
  • Computer measurements were consistently higher, suggesting detection of subtle ischemia missed by humans.

Conclusions:

  • Manual ST deviation measurements demonstrate high interrater reliability.
  • Computerized ECG analysis offers superior precision and may identify ischemia missed by manual review.
  • Researchers should avoid mixing computer and manual ST deviation scores in studies.