Related Experiment Videos
Quality control for DNA contamination in laboratories using PCR-based class II HLA typing methods
J M McCormack1, M L Sherman, D H Maurer
1Department of Pathology and Laboratory Medicine, Emory University School of Medicine, Atlanta, Georgia 30322, USA.
Human Immunology
|April 15, 1997
Summary
This study introduces a new primer set (RBQBf/RBQBr) for detecting DNA contamination in molecular histocompatibility typing. Optimizing wipe test sample concentrations is crucial for accurate polymerase chain reaction (PCR) results.
Area of Science:
- Molecular Biology
- Immunogenetics
- Laboratory Science
Background:
- Quality control in molecular histocompatibility typing is essential.
- Polymerase chain reaction (PCR) is commonly used to detect DNA contamination.
- Standardized methods are needed for reliable contamination detection.
Purpose of the Study:
- To develop and validate a primer set for detecting DNA contamination in histocompatibility testing.
- To assess the impact of sample concentration on PCR-based wipe tests.
- To optimize PCR protocols for enhanced sensitivity and accuracy.
Main Methods:
- Design and synthesis of a specific oligonucleotide primer set (RBQBf/RBQBr).
- Detection of genomic DNA and amplicon from reference cell lines and typing products.
- Validation studies using controlled DNA contamination and varying wipe test sample concentrations.
Main Results:
- The RBQBf/RBQBr primer set successfully detected genomic DNA (50-100 pg) and amplicon (1 copy).
- Detection of DR-B, DQ-B, and DP-B sequences was confirmed.
- Lower concentrations (1-2%) or diluted wipe test samples enhanced PCR sensitivity and optimized results, while higher concentrations (5-20%) were inhibitory.
Conclusions:
- The RBQBf/RBQBr primer set is effective for detecting DNA contamination and amplicon in molecular histocompatibility typing.
- Optimizing wipe test sample dilution is critical for sensitive and accurate PCR-based contamination detection.
- Routine testing for inhibitory factors in wipe tests is necessary for reliable laboratory quality control.