Interference with cardiac pacemakers by cellular telephones
D L Hayes1, P J Wang, D W Reynolds
1Mayo Clinic, Rochester, MN 55905, USA.
Background:
A growing body of evidence suggests that electromagnetic interference may occur between cardiac pacemakers and wireless hand-held (cellular) telephones, posing a potential public health problem. Electromagnetic interference may occur when the pacemaker is exposed to an electromagnetic field generated by the cellular telephone.
Methods:
In this multicenter, prospective, crossover study, we tested 980 patients with cardiac pacemakers with five types of telephones (one analogue and four digital) to assess the potential for interference. Telephones were tested in a test mode and were programmed to transmit at the maximal power, simulating the worst-case scenario; in addition, one telephone was tested during actual transmission to simulate actual use. Patients were electrocardiographically monitored while the telephones were tested at the ipsilateral ear and in a series of maneuvers directly over the pacemaker. Interference was classified according to the type and clinical significance of the effect.
Results:
The incidence of any type of interference was 20 percent in the 5533 tests, and the incidence of symptoms was 7.2 percent. The incidence of clinically significant interference was 6.6 percent. There was no clinically significant interference when the telephone was placed in the normal position over the ear. Interference that was definitely clinically significant occurred in only 1.7 percent of tests, and only when the telephone was held over the pacemaker. Interference was more frequent with dual-chamber pacemakers (25.3 percent) than with single-chamber pacemakers (6.8 percent, P<0.001) and more frequent with pacemakers without feed-through filters (28.9 to 55.8 percent) than with those with such filters (0.4 to 0.8 percent, P=0.01).
Conclusions:
Cellular telephones can interfere with the function of implanted cardiac pacemakers. However, when telephones are placed over the ear, the normal position, this interference does not pose a health risk.
Related Concept Videos
Electrophysiology of Normal Cardiac Rhythm
Mechanism of Cardiac Arrhythmias
Disturbances in Heart Rhythm
Arrhythmias are categorized by their speed, rhythm, and origin. A slow heart...
Cardiac Action Potential
The cardiac action potential process involves a series of phases characterized by the movement of ions across the cardiac cell membranes, leading to the depolarization and repolarization of the cardiac myocytes.
Ionic Basis of Cardiac Action Potentials
Holter Monitor: 24-Hour Monitoring
ECG Interpretation of Arrhythmias II: Atrial, Junctional and Ventricular Arrhythmias