Orthovoltage applicator design: its effect on a transmission monitor chamber
1Medical Physics Department, Addenbrooke's Hospital, Cambridge, UK.
Abstract:
In the Pantak DXT300, the applicators are mounted immediately below a transmission monitor chamber. Changing the metals used in applicator construction can cause variations of up to 22% in calibration factor (cGy MU(-1)) and apparent dose rate (MU min(-1)), but has no effect on radiation dose rate (cGy min(-1)). The results are consistent with being caused by backscatter from the applicator to the monitor chamber.