Memory artifact related to selenium-based digital radiography systems
H G Chotas1, C E Floyd, C E Ravin
1Department of Radiology, Duke University Medical Center, Durham, NC 27710, USA.
Abstract:
Digital images acquired on radiography systems with amorphous selenium detectors are susceptible to "memory artifacts" from prior x-ray exposures. In routine clinical use and in a laboratory experiment, artifacts appeared in chest radiographs until the selenium recovered from initial exposure. Memory artifacts were eliminated when 3 minutes or more elapsed between acquisition of a lateral chest radiograph and acquisition of the next radiograph.