Memory artifact related to selenium-based digital radiography systems

H G Chotas1, C E Floyd, C E Ravin

  • 1Department of Radiology, Duke University Medical Center, Durham, NC 27710, USA.

Radiology
|June 1, 1997
PubMed
Summary
This summary is machine-generated.

Related Concept Videos

You might also read

Related Articles

Articles linked to this work by shared authors, journal, and citation graph.

Sort by
Same author

Feature and knowledge based analysis for reduction of false positives in the computerized detection of masses in screening mammography.

Conference proceedings : ... Annual International Conference of the IEEE Engineering in Medicine and Biology Society. IEEE Engineering in Medicine and Biology Society. Annual Conference·2007
Same author

Application of the mutual information criterion for feature selection in computer-aided diagnosis.

Medical physics·2002
Same author

Computerized classification of suspicious regions in chest radiographs using subregion Hotelling observers.

Medical physics·2002
Same author

A neural network approach to breast cancer diagnosis as a constraint satisfaction problem.

Medical physics·2001
Same author

Digital chest radiography with a solid-state flat-panel x-ray detector: contrast-detail evaluation with processed images printed on film hard copy.

Radiology·2001
Same author

Imaging characteristics of an amorphous silicon flat-panel detector for digital chest radiography.

Radiology·2001

Digital radiography memory artifacts in amorphous selenium detectors can affect chest X-rays. Waiting at least three minutes between exposures eliminates these artifacts, ensuring clearer images.

Area of Science:

  • Medical Imaging Technology
  • Radiography Systems
  • Detector Physics

Background:

  • Digital radiography systems utilizing amorphous selenium detectors are prone to "memory artifacts" resulting from previous X-ray exposures.
  • These artifacts manifest as residual image data impacting the quality of subsequent radiographs.
  • Understanding the nature of these artifacts is crucial for maintaining diagnostic accuracy in clinical settings.

Purpose of the Study:

  • To investigate the occurrence and characteristics of memory artifacts in amorphous selenium digital radiography systems.
  • To determine the conditions under which these artifacts appear and persist.
  • To identify effective strategies for mitigating or eliminating memory artifacts in clinical practice.

Main Methods:

  • The study involved both routine clinical use of radiography systems and a controlled laboratory experiment.

Related Experiment Videos

  • Chest radiographs were acquired sequentially to observe the development and resolution of artifacts.
  • Varied time intervals between sequential X-ray exposures were analyzed to assess their impact on artifact persistence.
  • Main Results:

    • Memory artifacts were observed in chest radiographs acquired with amorphous selenium detectors, particularly following initial X-ray exposures.
    • These artifacts persisted until the selenium detector material sufficiently recovered from the prior exposure.
    • A minimum interval of three minutes between the acquisition of a lateral chest radiograph and the subsequent radiograph was found to completely eliminate memory artifacts.

    Conclusions:

    • Memory artifacts are a significant consideration for digital radiography systems employing amorphous selenium detectors.
    • Implementing a minimum three-minute delay between sequential radiographic exposures effectively resolves these artifacts.
    • This finding provides a practical guideline for optimizing image quality and diagnostic reliability in clinical radiography.