A simple method to measure compartment pressures using an intravenous catheter
Related Experiment Video
Updated: Nov 1, 2026

A Novel Non-invasive Method for the Detection of Elevated Intra-compartmental Pressures of the Leg
Published on: May 31, 2019
S C Wilson1, M S Vrahas, L Berson
1Louisiana State University Medical Center, New Orleans, USA.
Abstract:
A simple method was investigated to measure compartment pressures using 16-ga intravenous catheters with or without side ports attached by arterial line tubing to a pressure transducer. Pressure measurements from the experimental catheters were within 4 mm Hg of the slit catheter for 99% of all readings, and pressure measurements from the Stryker device were within 5 mm Hg of the slit catheter for 95% of all readings. The addition of one or two side ports to the experimental catheters did not alter the pressure readings. This method is comparable in accuracy to the slit catheter and in simplicity to the Stryker device.
Related Concept Videos
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Assessment of blood pressure in brachial artery(two-step method)
Sites for measuring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Cardiac Catheterization II: Right Heart Catheterization