Related Experiment Video
Updated: Aug 11, 2026

Use of a Hanging-weight System for Isolated Renal Artery Occlusion
Published on: July 19, 2011
[Experimental in vitro model of an artificial kidney--measurement of trace element levels using the ICP method.
P Koniński1, W Zyrnicki, A Steciwko
1Katedry i Kliniki Nefrologii Akademii Medycznej we Wrocławiu Kierownik.
Abstract:
Trace elements in blood serum is significantly changed in hemodialysed patients. The result of our experiment showed the manner and directions of the evaluated trace elements -Sr, Zn, Ni, Ba, B, Si, Mn and Cu movements. It was confirmed that reverse osmosis is efficient in the most tracers elimination, except Si, B, and Cu. The trace elements being present in the whole arrangement diffuse easily through the dialysers membrane used in our experiment, except Ni. In the membrane of the new, non-rinsed dialyser we detected Pb, Fe, V, Be, Ti and Al. It could possibly be the source of their contamination into the human organism. ICP-AES method of trace elements analysis reveals the possibilities of precisely, fast and simultaneous evaluation of big number various trace elements.