Related Experiment Video
Updated: Aug 14, 2026

06:02
2-Methacryloyloxyethyl Phosphorylcholine Polymer Treatment of Complete Dentures to Inhibit Denture Plaque Deposition
Published on: December 26, 2016
Development of a recording system for saliva pH with and complete denture by telemetry
T Watanabe1, K Kobayashi, T Suzuki
1Department of Geriatric Dentistry, Faculty of Dentistry, Tokyo Medical and Dental University, Japan. watanabe.gero@dent.tmd.ac.jp
Archives of Oral Biology
|March 1, 1997
Abstract:
A new system for measuring saliva pH by radio telemetry was developed. The system was built into a lower complete denture and could operate for about 19 h. The maximum error in the range pH 5-8 was 0.15 pH, and the correlation when values were compared with those from a glass-electrode pH meter was 0.99. That this system operates successfully was confirmed in a clinical experiment.