Virus retention by a hydrophilic triple-layer PVDF microporous membrane filter

H Aranha-Creado1, K Oshima, S Jafari

  • 1Pall Corporation, Port Washington, New York, USA.

Insights

This study shows the Ultipor VF DV50 membrane effectively removes viruses larger than 50 nm, achieving a minimum log titer reduction of 6. It also maintains high protein transmission, making it suitable for biopharmaceutical manufacturing.

Area of Science:

  • Biotechnology
  • Filtration Technology
  • Virology

Background:

  • Effective viral clearance is critical in biopharmaceutical manufacturing to ensure product safety.
  • Microporous membranes are widely used for virus removal, but performance can vary with carrier fluid composition.
  • Hydrophilic polyvinylidene fluoride (PVDF) membranes offer potential for high protein transmission and viral retention.

Purpose of the Study:

  • To evaluate the viral retention capabilities of the Ultipor VF grade DV50 hydrophilic triple-layer PVDF microporous membrane.
  • To assess the impact of various carrier fluids on virus retention and protein transmission.
  • To demonstrate the correlation between a nondestructive integrity test and virus retention in filter cartridges.

Main Methods:

  • Virus retention studies were conducted using bacteriophages (phi 6, PR772, T1, PP7) and mammalian viruses (poliovirus, influenza A virus).
  • Filtration challenges were performed on membrane discs and pleated filter cartridges at viral concentrations of 10(6)-10(8) PFU/mL.
  • Carrier fluids included water, saline, gelatin buffer, MEM, and MEM with 10% fetal bovine serum.
  • Protein transmission for IgG and albumin was measured.
  • Forward flow integrity testing was used to correlate with virus retention in pleated cartridges.

Main Results:

  • A minimum log titer reduction (LTR) of 6 was consistently achieved for viruses larger than 50 nm, regardless of the carrier fluid.
  • Protein transmission levels exceeded 95% for both IgG and albumin.
  • A strong correlation was demonstrated between the forward flow integrity test and virus retention performance in pleated filter cartridges.

Conclusions:

  • The Ultipor VF grade DV50 membrane demonstrates robust viral retention (>6-log reduction) and high protein transmission (>95%).
  • The filter's performance is consistent across various biologically relevant carrier fluids.
  • The demonstrated correlation between integrity testing and viral retention supports its use in quality control for biopharmaceutical manufacturing.