Dosimeter gel and MR imaging for verification of calculated dose distributions in clinical radiation therapy
Related Experiment Video
Updated: Aug 16, 2026

Dosimetry for Cell Irradiation using Orthovoltage (40-300 kV) X-Ray Facilities
Published on: February 20, 2021
S A Johansson1, P Magnusson, A Fransson
1Department of Radiation Physics, Malmö, Lund University, Malmö University Hospital, Sweden. Sven.Johansson@rfa.mas.lu.se
Abstract:
A dosimeter gel, based on an agarose gel infused with a ferrous sulphate solution and evaluated in a magnetic resonance scanner, was used for complete verification of calculated dose distributions. Two standard treatment procedures, treatment of cancer in the urinary bladder and treatment of breast cancer after modified radical mastectomy, were examined using pixel-by-pixel and dose volume histogram comparison. The dose distributions calculated with the dose planning system was in very good agreement with the measured ones. However, in the case of the more complicated breast cancer treatment, some discrepancies were found, mainly at the beam abutment region. This may be explained by field displacements errors and by a small limitation of the dose planning utilising small electron beams in this region. The dosimeter gel system have proven to be a useful tool for dosimetry in clinical radiation therapy applications.
More Related Videos
06:20Irradiator Commissioning and Dosimetry for Assessment of LQ α and β Parameters, Radiation Dosing Schema, and in vivo Dose Deposition
Published on: March 11, 2021
07:57Positron Emission Tomography-based Dose Painting Radiation Therapy in a Glioblastoma Rat Model using the Small Animal Radiation Research Platform
Published on: March 24, 2022
Related Concept Videos
Biological Effects of Radiation
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET