Impact of setup variability on incidental lung irradiation during tangential breast treatment
D L Carter1, L B Marks, G C Bentel
1Department of Radiation Oncology, Duke University Medical Center, Durham, NC 27710, USA.
Purpose:
This study aimed to determine the variability in treatment setup during a 5-week course of tangential breast treatment for patients immobilized in a customized hemibody cradle, to assess the relationship between the height of the lung shadow on the tangential port film and the percentage of lung volume irradiated, and to estimate the impact of setup variabilities on irradiated lung volume.
Methods:
One hundred seventy-two port films were reviewed from 20 patients who received tangential beam treatment for breast cancer. The height of the lung shadow at the central axis (CLD) on each port film was compared to the corresponding simulator film as an assessment of setup variability. A three-dimensional dose calculation was performed, and the percentage of total lung volume within the field was correlated with the CLD. The three-dimensional dose calculation was repeated for selected patients with the location of the treatment beams modified to reflect typical setup variations.
Results:
The CLD measured on the port films was within 3 mm of that prescribed on the simulator film in 43% (74 of 172) of the port films. The variation was 3-5 mm in 26%, 5-10 mm in 25%, and >10 mm in 6%. The height of the lung shadow correlated with the percentage of lung volume included in the radiation field (r2 = 0.6). Typical variations in treatment setup resulted in < or = 5% fluctuation in the absolute volume of ipsilateral lung irradiated.
Conclusion:
The current immobilization system used in our clinic provides a clinically acceptable reproducibility of patient setup. The height of the lung shadow is reasonably well correlated with the percentage of irradiated lung volume. During a typical 5-week course of radiotherapy, the ipsilateral irradiated lung volume fluctuates <5%.