Related Experiment Videos
Evaluation of hemolysis in a pulsatile assist device for centrifugal pump
1Department of Cardiovascular Surgery, Fukushima Medical College, Japan.
Abstract:
To evaluate the blood trauma caused by a new device for producing a pulsatile flow of the centrifugal pump, the pulsatile assist device for the centrifugal pump (PAD-CP) that we have developed, a hemolysis study was performed in vitro and in animal experimentation. For the in vitro testing, 2 identical sets of hemolysis test circuits were prepared with 2,400 ml of bovine blood. The 2 circuits were pumped simultaneously. Plasma total hemoglobin levels were less than 40 mg/dl after 3 h, under a pump flow of 2 L/min. Hemolysis increased to a severe level after 4 h of 4 L/min pump flow. The cause of this hemolysis was thought to be a vibration of the circuit because of incomplete compression of the polyurethane tube in the PAD-CP. Five adult sheep (average body weight, 47 kg) were used for in vivo evaluation of hemolysis. Hemolysis was less than 30 mg/dl of plasma hemoglobin after 4 h of open chest extracorporeal circulation with 3.0-3.6 L/min of flow rate using the PAD-CP. Other hematologic changes after PAD-CP driving were within normal limits. We conclude that the PAD-CP has proven to have possible clinical applications.