Spectral characterization and classification of Carpentier-Edwards heart valves implanted in the aortic position
Related Experiment Video
Updated: Nov 1, 2026

Echocardiographic Approaches and Protocols for Comprehensive Phenotypic Characterization of Valvular Heart Disease in Mice
Published on: February 14, 2017
H P Sava1, P M Grant, J T McDonnell
1Department of Electrical Engineering, University of Edinburgh, U.K. herkole@essex.ac.uk
Abstract:
This paper demonstrates an improvement in the performance of spectral phonocardiography, combined with pattern recognition techniques for monitoring the condition of bioprosthetic heart valves. The analysis of the heart sounds is performed using a modified forward-backward overdetermined Prony's method. Results show that the condition of the bioprosthesis affects mostly the higher part of the spectrum (i.e., above 250 Hz) where no frequency components were found for malfunctioning cases. Therefore, the amplitudes of the three highest frequency components are used as the input vector of an adaptive single layer perceptron-based classifier to identify normal and malfunctioning classes. For the sample set examined, this method gives 100% correct discrimination between normal and malfunctioning Carpentier-Edwards (C-E) valves.
Related Concept Videos
Aortic Regurgitation II: Clinical Features and Diagnostic Tests
Mitral Stenosis II: Clinical features and Diagnostic Tests
Aortic Regurgitation I: Introduction
Heart Valves
The AV valves prevent the backflow of blood from the ventricles to the atria during ventricular contraction. These valves function with the assistance of the chordae tendineae and papillary muscles. When the ventricles are relaxed, the chordae tendineae are slack, allowing blood to flow from the atria into the...