Quality control programme in mammography: second level quality controls

E Nassivera1, L Nardin

  • 1Department of Medical Physics, S. Chiara Hospital, Trento, Italy.

The British Journal of Radiology
|June 1, 1997
PubMed
Summary

Regular quality control ensures mammography equipment functions correctly for reliable breast lesion detection. Comprehensive checks verify image quality and radiation dose, essential for diagnostic accuracy and patient safety.

Related Concept Videos

You might also read

Related Articles

Articles linked to this work by shared authors, journal, and citation graph.

Sort by
Same author

Daily quality control programme in mammography.

The British journal of radiology·1996
Same author

[Radiotherapy of T1N0 neoplasms of the glottis. Analysis of the parameters that influence local control].

La Radiologia medica·1995
Same author

[Quality assurance of mammography in the province of Trent].

La Radiologia medica·1993
Same author

[Sotalol in the Wolff-Parkinson-White syndrome: an electrophysiological and clinical study].

Giornale italiano di cardiologia·1992
Same author

[Radiotherapy, radiotherapy combined with surgery, radiotherapy combined with chemotherapy in the treatment of carcinoma of the pyriform sinus. A retrospective study].

La Radiologia medica·1991
Same author

[Radical radiation therapy of carcinoma of the rhinopharynx: analysis of 80 cases].

La Radiologia medica·1990

Area of Science:

  • Radiology
  • Medical Physics
  • Diagnostic Imaging

Background:

  • Mammography is crucial for breast lesion detection, but low contrast requires high image quality.
  • Maintaining high mammographic standards necessitates stringent quality control to ensure diagnostic accuracy.

Purpose of the Study:

  • To describe the second level of a mammography quality control program.
  • To detail the parameters and methods used for semi-annual equipment checks by medical physicists.

Main Methods:

  • Implemented a two-level quality control program with daily and semi-annual checks.
  • Conducted detailed semi-annual assessments of mammographic units, screen-film systems, and radiation dose.
  • Measured parameters including kilovoltage, focal spot size, half-value layer, tube output, AEC, and mean glandular dose.

Related Experiment Videos

Main Results:

  • Evaluated key mammographic parameters over a five-year period (1991-1995).
  • Results demonstrated consistent and correct functioning of the mammography equipment throughout the study.
  • Ensured adherence to national and international quality standards for mammography.

Conclusions:

  • The described quality control program effectively verifies mammographic equipment performance.
  • Consistent equipment function is vital for accurate breast lesion detection and patient safety.
  • Semi-annual assessments by medical physicists are essential for maintaining high mammography standards.