Related Experiment Videos
A specific and sensitive ELISA for measuring S-100b in cerebrospinal fluid
A J Green1, G Keir, E J Thompson
1Department of Neuroimmunology, National Hospital for Neurology and Neurosurgery, London, UK. skgtejt@ion.bpmf.ac.uk
Journal of Immunological Methods
|June 23, 1997
Summary
A new enzyme-linked immunosorbent assay (ELISA) for S-100b protein in cerebrospinal fluid (CSF) is sensitive and specific. This assay can detect elevated S-100b levels in neurological conditions like cerebral hemorrhage and CNS malignancy.
Area of Science:
- Biochemistry
- Immunology
- Clinical Chemistry
Background:
- S-100b is a protein found in cerebrospinal fluid (CSF).
- Elevated S-100b levels are associated with various neurological conditions.
- A sensitive and specific assay is needed for accurate S-100b quantification in CSF.
Purpose of the Study:
- To develop and validate a sensitive, simple, and specific sandwich ELISA for quantifying S-100b in human CSF.
- To assess the diagnostic utility of CSF S-100b levels in patients with neurological diseases.
Main Methods:
- A sandwich ELISA was developed using a monoclonal anti-S-100b capture antibody and a horseradish peroxidase-labeled polyclonal anti-S-100b detector antibody.
- The assay was validated for sensitivity, specificity, precision, and recovery using S-100b standards and spiked CSF samples.
- 154 CSF samples from patients with various neurological conditions and controls were analyzed.
Main Results:
- The ELISA demonstrated a lower limit of detection of 0.04 ng/ml with minimal cross-reactivity (<0.006%) with S-100a.
- The assay showed good precision (within-batch CV 5.6-9.3%, between-batch CV 8.1-8.9%) and recovery (94%).
- Elevated CSF S-100b levels (>0.4 ng/ml) were found in 19% of samples, with the highest concentrations in patients with cerebral hemorrhage or CNS malignancy.
Conclusions:
- The developed sandwich ELISA is a reliable and efficient method for routine S-100b measurement in clinical laboratories.
- CSF S-100b levels can serve as a potential biomarker for detecting and monitoring neurological diseases, particularly cerebral hemorrhage and CNS malignancy.
- S-100b appears to be rapidly cleared from CSF, as evidenced by a dramatic decline in levels in a patient with an infected shunt.