Dynamic Lung Tumor Tracking for Stereotactic Ablative Body Radiation Therapy
Published on: June 7, 2015
Phantom assessment of lung dose from proton arc therapy
G A Sandison1, E Papiez, C Bloch
1Department of Radiation Oncology, Indiana University Medical Center, Indianapolis, USA.
Purpose:
To compare resultant lung dose from proton arc therapy of the chest wall to that from electron arc therapy.
Methods And Materials:
A 200 MeV proton beam from the Indiana University Cyclotron was range shifted and modulated to provide a spread out Bragg peak extending from the surface to a depth of 4 cm in water. The chest wall of an Alderson Rando phantom was irradiated by this beam, collimated to a 20 x 4 cm field size, while it rotated on a platform at approximately 1 rpm. For comparison, electron arc therapy of the Rando phantom chest wall was similarly performed with 12 MeV electrons and the resultant lung dose measured in each case.
Results:
Dose-volume histograms for the Rando phantom left lung indicate a reduced volume of irradiated lung for protons at all dose levels and an integral lung dose that is half that for electron arc therapy in the case studied. In addition, a more uniform dose coverage of the target volume was achieved with the proton therapy.
Conclusion:
This study demonstrates a potential role for proton arc therapy as an alternative to electron arc therapy when lung dose must be minimized.