CT arterial portography and CT arteriography with a triple-lumen balloon catheter

T Murakami1, H Oi, M Hori

  • 1Department of Radiology, Osaka University Medical School, Suita-City, Japan.

Insights

The triple-lumen balloon catheter is a useful tool for performing both CT arterial portography (CT-AP) and CT arteriography (CT-A) sequentially. This technique proved convenient and safe in a study of 58 patients.

Area of Science:

  • Interventional Radiology
  • Medical Imaging

Background:

  • CT arterial portography (CT-AP) and CT arteriography (CT-A) are valuable diagnostic imaging techniques.
  • Serial performance of these procedures can enhance diagnostic yield.
  • A specialized catheter may improve the efficiency and safety of combined examinations.

Purpose of the Study:

  • To assess the utility of a triple-lumen balloon catheter for serial CT-AP and CT-A procedures.
  • To determine the feasibility and effectiveness of this catheter in a clinical setting.

Main Methods:

  • A cobra-type triple-lumen balloon catheter was used in 58 patients for combined CT-AP and CT-A.
  • Contrast administration techniques varied for CT-AP (splenic or superior mesenteric artery) and CT-A (hepatic or accessory hepatic arteries).
  • Catheter placement and contrast delivery were optimized using side-holes and end-holes relative to the inflated balloon.

Main Results:

  • Successful CT-AP was achieved in 91% (53/58) of patients.
  • Successful CT-A was achieved in 72% (42/58) of patients.
  • Combined successful CT-AP and CT-A were obtained in 72% (42/58) of patients, with no reported complications.

Conclusions:

  • The triple-lumen balloon catheter facilitates convenient and effective serial CT-AP and CT-A.
  • This technique offers a safe approach for combined arterial imaging.
  • The catheter's design supports versatile contrast delivery for comprehensive vascular assessment.