Stimulus artefact in somatosensory evoked potential measurement
Related Experiment Video
Updated: Nov 1, 2026

Somatosensory Event-related Potentials from Orofacial Skin Stretch Stimulation
Published on: December 18, 2015
R N Scott1, L McLean, P A Parker
1Institute of Biomedical Engineering, University of New Brunswick, Fredericton, Canada.
Abstract:
When an electrical stimulus is used to evoke action potentials in peripheral nerves or the spinal cord, the stimulus causes an artefact which may interfere with measurement of the evoked potentials. This artefact, unlike all other sources of noise in the measurement, cannot be reduced by ensemble averaging. Confusion about the origin and transmission of stimulus artefact has led to considerable frustration in spinal somatosensory evoked potential (SSEP) measurements. The three mechanisms by which stimulus artefact is coupled into the measuring system are identified, and means for their reduction are discussed.