Related Experiment Videos

Evaluation of i-STAT portable clinical analyzer in a neonatal and pediatric intensive care unit

J N Murthy1, J M Hicks, S J Soldin

  • 1Department of Laboratory Medicine, Children's National Medical Center, Washington, DC 20010, USA.

Insights

The Point-of-Care (POC) i-STAT system accurately measures blood gases and electrolytes in neonatal and pediatric intensive care units. This reliable analyzer offers a marginal improvement in turnaround time compared to traditional laboratory methods.

Area of Science:

  • Clinical Chemistry
  • Point-of-Care Testing
  • Critical Care Diagnostics

Background:

  • Traditional blood gas analysis requires laboratory processing, leading to potential delays in critical care settings.
  • Point-of-care testing (POCT) offers rapid results at the patient's bedside, improving clinical decision-making.
  • The i-STAT system is a handheld POCT device for various blood analyses.

Purpose of the Study:

  • To evaluate the accuracy and reliability of the i-STAT system for blood gas and electrolyte measurements in neonatal and pediatric intensive care units (NICU/PICU).
  • To assess the performance of the i-STAT system when operated by non-laboratory personnel.
  • To determine the necessity of routine quality control (QC) for the i-STAT system in a clinical setting.

Main Methods:

  • The i-STAT system's imprecision and linearity were assessed using aqueous standards and whole blood samples.
  • Patient samples were analyzed using both the i-STAT system and a Ciba Coming 288 blood gas analyzer for comparison.
  • The impact of operator (nurses vs. lab personnel) and routine QC on i-STAT results was evaluated.

Main Results:

  • The i-STAT system demonstrated good precision (%CVs < 3.5) and linearity for pH, pCO2, Na+, K+, and iCa2+.
  • Excellent correlation (r > 0.95) was observed between i-STAT and Ciba Coming 288 for pH, pCO2, pO2, Na+, and K+; iCa2+ showed moderate correlation (r = 0.73).
  • No significant difference in results was found between NICU/PICU nurses and laboratory personnel, and routine QC was deemed unnecessary.

Conclusions:

  • The i-STAT system is a reliable point-of-care alternative to traditional blood gas analyzers for neonatal and pediatric critical care.
  • The system provides a marginal improvement in turnaround time compared to central laboratory services.
  • Careful cost management is essential for the effective implementation of the i-STAT system.
Abstract

Related Concept Videos