Determination of silicon in serum and tissue by electrothermal atomic absorption spectrometry
1Department of Clinical Biochemistry, London Health Sciences Centre, Ontario, Canada. fleung@julian.uwo.ca
Objectives:
To employ an electrothermal atomic absorption spectrometry (ETAAS) method with chemical modifiers for the determination of silicon (Si) in serum and tissues.
Methods:
Si was measured in serum of chronic hemodialysis patients, and in fibrous breast capsule tissues following silicone-gel implant removal. Tissue was dried, and digested with concentrated nitric acid prior to analysis. A chemical modifier, which included lanthanum oxide and ammonium phosphate, was used to dilute the serum, and digested tissue samples (1:4) before atomization. Si was determined at 251.6 nm in a graphite fumace using an atomic absorption spectrophotometer.
Results:
The method was linear (to 1000 micrograms/L), and precise (CV 5.9% at 281 micrograms/L and 8.4% at 73 micrograms/L). Recovery of Si in spiked serum and breast tissue samples was between 97-104%. Reference values for women on a normal diet and no exposure to silicone implants gave serum Si to 30 to 209 micrograms/L (n = 60), and breast tissue levels of 0.25 to 2.4 micrograms/g dry wt (n = 48). Si in breast capsule from women exposed to silicone breast implants varied from 29 to 496 micrograms/g dry wt (n = 10). Serum Si in hemodialysis patients (n = 53) ranged from 900 to 3300 micrograms/L.
Conclusions:
We conclude that our chemically modified ETAAS method is suitable for Si determination in normal and elevated human serum and tissue specimens.
Related Concept Videos
Atomic Absorption Spectroscopy: Overview
When irradiated by EMR of a particular wavelength, these...
Atomic Absorption Spectroscopy: Lab
Solutions containing organic solvents, such as low-molecular-mass alcohols, esters, or ketones, enhance absorbances by increasing nebulizer...
Atomic Emission Spectroscopy: Overview
Atomic Emission Spectroscopy: Instrumentation
Inductively Coupled Plasma Atomic Emission Spectroscopy: Principle
The ions and electrons produced interact with the fluctuating magnetic field created by a water-cooled...
Atomic Emission Spectroscopy: Lab