Implantable cardioverter defibrillator oversensing during periods of rate-related bundle branch block
R F Coyne1, D C Man, B H Sarter
1Allegheny University of the Health Sciences, Philadelphia, Pennsylvania 19129, USA.
Journal of Cardiovascular Electrophysiology
|July 1, 1997
Abstract:
A patient with an implantable cardioverter defibrillator (ICD) and a dual chamber pacemaker experienced inappropriate ICD therapies only during periods of rate-dependent right bundle branch block. Analysis of both stored and real-time ICD electrograms was critical to correctly diagnosing the problem and offering a solution.