Related Experiment Videos
Coulter counting and light diffraction analysis applied to characterisation of oil-water emulsions
P C Sontum1, E M Kolderup, D in't Veldt
1Analytical Sciences R&D, Nycomed Imaging AS, Oslo, Norway. Per.sontum@nycomed.telemax.no
Abstract:
Coulter counting and light diffraction techniques were successfully applied to the characterisation of the droplet concentration and size distribution in camphene-water and cyclohexane-water emulsions. Both of these techniques required a dilution of the emulsion prior to analysis, and it was found that the destabilizing effect on the droplets of such dilution could be overcome by performing the analyses at temperatures below the melting point of the oil phase. The storage stability of the camphene-water samples at 60 degrees C was reasonably good with a 5-20% change in the investigated parameters over a 24 h period. At room temperature camphene-water samples left to stand undisturbed were unaffected after 24 h, while continuous mixing of the emulsion on a roller board brought about a rapid amalgamation of the particles into larger aggregates. This fusion process was fully described only with the light diffraction analysis due to the broader measuring range of this technique. However, analysing emulsions with a droplet size range covered by both techniques gave identical results.