Behavior of nitric oxide infused at constant flow rates directly into a breathing circuit during controlled
J W Skimming1, P B Blanch, M J Banner
1Department of Pediatrics, University of Florida College of Medicine, JHMHC, Gainesville 32610-0296, USA.
Objectives:
This study was designed to test the hypothesis that the practice of infusing nitric oxide at constant flow rates directly into breathing circuits with intermittent (pulsatile) flow can lead to streaming and tidal pooling of the nitric oxide. This study was also designed to show the extent to which streaming and tidal pooling of nitric oxide affect nitric oxide delivery.
Design:
A series of five in vitro experiments was performed. For each experiment, either one or two features of the nitric oxide delivery/sampling system were varied, and the effects of these variations were evaluated with regard to measured nitric oxide concentration changes. The results from each experiment were analyzed using either one- or two-factor analysis of variance.
Setting:
University research laboratory.
Subjects:
Breaths were provided by a mechanical ventilator that was connected to a lung model. A standard, corrugated, adult breathing circuit was used. Gas samples were obtained from either the lung model's bellows or selected sites within the breathing circuit. Nitric oxide concentrations were measured, using an electrochemical gas analyzer.
Interventions:
The system features that were varied included the cross-sectional position of the sampling site within the breathing circuit, the distance between the infusion port and the sampling site, the breathing frequency, the distance between the Y-piece and the infusion port, and the airway (deadspace) volume.
Measurements And Main Results:
Streaming of nitric oxide within the breathing circuit was detected as far as 25 cm downstream of the infusion site (p < .0001). Pooling of nitric oxide was detected both near and downstream of the infusion site (p < .0001). Increasing the breathing frequency from 5 to 30 breaths/min increased mixing thoroughness (p < .005). Increasing the distance between the Y-piece and the infusion port from 15 to 180 cm decreased nitric oxide delivery to our lung model (p < .0001). Interestingly, increasing airway (deadspace) volume from 150 to 450 mL decreased nitric oxide delivery to our lung model (p < .0001).
Conclusions:
Estimates of nitric oxide delivery using a constant flow rate of nitric oxide infused directly into a breathing circuit during controlled mechanical ventilation can be confounded by streaming and tidal propagation of nitric oxide pools. Improved reproducibility of reported dose-response relationships is likely to be achieved through further study of nitric oxide behavior within the breathing circuits. Reduced toxicity associated with nitric oxide inhalation may also be achieved through a better understanding of this nitric oxide behavior.
More Related Videos
08:35Oxygenation-sensitive Cardiac MRI with Vasoactive Breathing Maneuvers for the Non-invasive Assessment of Coronary Microvascular Dysfunction
Published on: August 17, 2022
07:15Preoxygenation Techniques for Tracheal Intubation in Critically Ill Adults Utilizing Oxygen Mask and Noninvasive Ventilation
Published on: December 5, 2025
Related Concept Videos
Physiological Control of Respiration
Breathing, a seemingly passive process, is regulated by the respiratory center in the brainstem. This center coordinates the involuntary control of respirations, which means it occurs without conscious effort, ensuring a smooth and uninterrupted pattern.
Regulation of Ventilation
The body maintains ventilation by monitoring levels of carbon dioxide (CO2), oxygen (O2), and hydrogen ion concentration (pH) in the arterial blood. Among these factors, the level of CO2 plays a crucial...
Oxygen Delivering System I: Nasal Cannula and Face Mask
Nasal Cannula
A nasal cannula is a lightweight tube split at one end into two prongs and placed in the nostrils. It is typically used to deliver low to medium levels of oxygen.
Suggested flow rate: The suggested flow rate for a nasal cannula typically ranges between 1 and 6 L/min.
Oxygen percentage setting:...
Mechanical Ventilation I: Indication and Settings
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...
Mechanical Ventilation III: Noninvasive Ventilation
Noninvasive Positive-Pressure Ventilation (NIPPV)
Ventilatory Modes
There are three ventilatory modes: full support, partial support, and spontaneous. These are described below.
Full Support Modes
Full support modes include controlled mechanical ventilation, continuous mandatory...