Related Experiment Videos

Percutaneous treatment of iliac aneurysms and pseudoaneurysms with Cragg Endopro System 1 stent-grafts

D Gasparini1, A Lovaria, S Saccheri

  • 1Institute of Radiology, Division of Vascular and Interventional Radiology, Azienda Ospedaliera S. Maria della Misericordia, Piazza S. Maria M, I-33100 Udine, Italy.

Insights

The Cragg Endopro System 1 stent-graft offers a safe and effective method for treating iliac artery aneurysms, demonstrating high success rates in short-term follow-up studies.

Area of Science:

  • Vascular Surgery
  • Interventional Radiology

Background:

  • Iliac artery aneurysms pose significant risks.
  • Endovascular repair offers a minimally invasive alternative to open surgery.

Purpose of the Study:

  • To assess the feasibility and short-term outcomes of treating iliac aneurysms using the Cragg Endopro System 1 stent-graft.

Main Methods:

  • Percutaneous implantation of 10 Cragg Endopro System 1 stent-grafts in 9 lesions (2 pseudoaneurysms, 7 atherosclerotic aneurysms) across 8 patients.
  • Post-procedure anticoagulation with heparin followed by antiplatelet therapy.
  • Follow-up included Doppler ultrasound at 2 days and 3 months, with angiography/angio-CT up to 12 months for some patients.

Main Results:

  • Achieved a 100% initial clinical success rate.
  • Reported two minor complications, including one device deployment failure requiring a second graft.
  • All prostheses remained patent at 3-12 months, with one instance of minimal perigraft leak (12.5%).

Conclusions:

  • Percutaneous stent-grafting with the Cragg Endopro System 1 is a safe and effective treatment for iliac artery aneurysms.
Abstract

Related Concept Videos