[Comparison of radiological procedures in preoperative assessment of anchorage strength of spinal implants]
Related Experiment Video
Updated: Aug 7, 2026

An Improved Mechanical Testing Method to Assess Bone-implant Anchorage
Published on: February 10, 2014
M Schwitalle1, P Eysel, A Eckardt
1Orthopädische Klinik, Johannes Gutenberg-Universität, Mainz.
Purpose:
To compare three techniques with regards to their ability to estimate pull out strength of spinal fusion hardware.
Material And Methods:
VDS-screw fixation strength in 50 human cadaveric vertebral bodies was approximated by means of pull out force measurement. Bone quality was assessed by Dual X-ray Absorptiometry (DXA). Quantitative Computed Tomography (QCT) and T2*-relaxation time. For each of these techniques, correlation with axial pull out force strength was investigated.
Results:
Highest correlation was found for cancellous bone density measured by QCT (r = 0.72; p < 0.001). Immediately followed by DXA (r = 0.70; p < 0.001), which involves all bone components. Inverted T2*-relaxation time (r = 0.55; p < 0.001) and cortical bone density (QCT) correlated just slightly with pull-out force strength.
Conclusions:
Absorptiometrical techniques like QCT and DXA are most appropriate to estimate VDS-screw fixation strength preoperatively.