Abstract
Objective:
The objective of our study was to assess the ability of a 0.5-T superconducting, open-configuration MR system to guide laser diskectomies. The study was performed ex vivo on six human cadavers as well as in vivo on three patients.
Conclusion:
The macroscopic size of necrosis correlated well with the monitored temperature spread (r2 = .76-.85). After MR-guided introduction of the laser fiber into the targeted disk space, temperature spread was visualized in all subjects.