Related Experiment Video
Updated: Jul 26, 2026

Human In-Vivo Bioassay for the Tissue-Specific Measurement of Nociceptive and Inflammatory Mediators
Published on: December 1, 2008
Pyrogenic reactions in patients undergoing cardiac catheterization associated with contaminated glass medicine cups
S T Cookson1, J J Nora, J A Kithas
1Hospital Infections Program, Centers for Disease Control and Prevention, Atlanta, Georgia 30333, USA.
Insights
Bacterial endotoxin in reused medicine cups caused life-threatening pyrogenic reactions in cardiac catheterization patients. Contaminated cups, linked to nitroglycerin exposure, led to severe illness and two deaths.
Area of Science:
- Infectious Disease
- Cardiology
- Patient Safety
Background:
- Pyrogenic reactions, a severe complication of bacterial endotoxin exposure, pose significant risks.
- Two cardiac catheterization patients experienced rigors, prompting an investigation into the cause.
Purpose of the Study:
- To identify the source of pyrogenic reactions in cardiac catheterization patients.
- To determine the association between specific procedural elements and patient rigors.
Main Methods:
- A case control study was conducted involving patients who developed rigors within 3 hours of cardiac catheterization.
- Medical records were reviewed to identify case patients and randomly selected control patients.
- Exposure to intracoronary-nitroglycerin (NTG) and the reprocessing of glass medicine cups were specifically examined.
Main Results:
- Twelve case patients and 40 control patients were identified.
- Exposure to intracoronary-nitroglycerin (NTG) was significantly associated with developing rigors (odds ratio = 12.0).
- Reprocessed glass medicine cups used for NTG demonstrated elevated endotoxin levels (median 2,250 EU/mL), originating from contaminated enzyme cleaner.
Conclusions:
- Contaminated, reprocessed glass medicine cups likely caused pyrogenic reactions in cardiac catheterization patients.
- Endotoxin exposure from these cups contributed to the death of two critically ill patients.
- This highlights the critical importance of sterile processing and endotoxin control in medical devices and supplies.
Abstract:
Pyrogenic reactions are potentially life-threatening complications caused by bacterial endotoxin. After two cardiac catheterization patients developed rigors the same day, the procedures were halted and a case control study was conducted. To identify case patients (persons with rigors < or = 3 hr after catheterization during September 25-November 9, 1995), we reviewed medical records of all cardiac catheterization patients who had a blood culture or received intravenous meperidine. Twelve case patients and 40 randomly selected control patients were identified. No specific catheter was associated with case patients, but exposure to intracoronary-nitroglycerin (NTG) was (odds ratio = 12.0; 95% confidence interval 2.2, 75.6). NTG or indocyanine green dye was poured into glass medicine cups previously washed in an enzyme cleaner and then sterilized. The cleaner, used for an entire day, had elevated levels of gram-negative bacteria (> 10(4) colony forming units/mL) and endotoxin (434 endotoxin units [EU]/mL]); the reprocessed cups had no live bacteria but had elevated endotoxin levels (median 2,250 EU). Exposure to contaminated glass medicine cups probably resulted in pyrogenic reactions and contributed to death in two critically ill patients.
Related Concept Videos
Hypersensitivity Reactions: Cytolytic Reactions
Cardiac Catheterization I: Pre-Procedure Overview
Cardiac Catheterization II: Right Heart Catheterization
Cardiac Catheterization IV: Nursing Management
Endocarditis I: Introduction
Hypersensitivity Reactions: Immune-Complex Reactions