Indwelling pericardial drainage catheter break secondary to heart movement and catheter angulation

C T Johnson1, L A Osborn

  • 1Department of Internal Medicine, University of New Mexico School of Medicine, Albuquerque. losborn@medusa.unm.edu

Insights

A pericardial drainage catheter broke during removal, necessitating surgery. Analysis revealed no defects, suggesting heart movement and insertion angle caused the catheter to sever.

Area of Science:

  • Cardiology
  • Medical Device Engineering
  • Surgical Complications

Background:

  • Pericardial drainage catheters are used to remove fluid from the pericardial space.
  • Catheter retrieval can occasionally be complicated by device failure.
  • Understanding failure mechanisms is crucial for patient safety.

Observation:

  • A case is presented where a pericardial drainage catheter fractured upon attempted removal.
  • Surgical intervention was required to retrieve the retained catheter fragment.
  • The removed catheter fragment was analyzed for potential defects.

Findings:

  • No manufacturing defects were identified in the retrieved catheter portion.
  • Suspected causes for the catheter disruption include shear forces from cardiac motion.
  • The angle of catheter entry into the pericardial space is also implicated.

Implications:

  • This case highlights potential risks associated with pericardial drainage catheter removal.
  • Understanding mechanical forces is crucial for preventing future catheter failures.
  • Further research into catheter design and insertion techniques may be warranted.