Related Experiment Videos
[Can digital selenium-based radiography in thoracic diagnosis replace the analog x-ray imaging technic?]
M Zähringer1, B Krug, W Dölken
1Institut und Poliklinik für Radiologische Diagnostik, Universität zu Köln.
Purpose:
To find out the diagnostic value of digital selenium radiography, we compared the image quality of chest x-ray images from 50 patients who had been examined via conventional chest x-ray and digital selenium radiography of the chest.
Method:
50 patients with a malignant melanoma underwent chest x-ray within 3 months in conventional technique and with digital selenium radiography (Thoravision: Philips Medical Systems, Hamburg, Germany). In this period none of the patients showed a difference in respect of clinical status or radiological diagnosis. Simultaneous examinations on the same day were not performed to avoid unnecessary exposure to x-rays. The digital and conventional images were compared by 4 radiologists with regard to image quality by the detection of defined anatomic structures.
Results:
Image quality of digital selenium radiography was considered superior to that of conventional chest x-rays in the mediastinum, the retrocardiac and retrodiaphragmatic areas, the superior and inferior lobes of the lung especially near the parietal pleura, and the chest wall.
Conclusions:
Compared to analogous techniques there is no loss of image information when employing digital selenium radiography in chest x-rays. On the contrary, new assessment criteria may be gained. We conclude that digital selenium radiography offers diagnostic advantages in chest x-ray examination.