Related Experiment Videos
Use of transtympanic round window electrocochleography for threshold estimations in children
S H Wong1, W P Gibson, H Sanli
1Department of Surgery, University of Hong Kong, Queen Mary Hospital, Hong Kong.
Objective:
This study aimed to evaluate the efficacy of round window electrocochleography for the estimation of hearing thresholds in difficult-to-test children.
Study Design:
The study was a retrospective analysis.
Setting:
A standard day-stay operating room was used.
Patients:
Round window electrocochleography was performed on 198 children between January 1993 and January 1996.
Intervention:
The intervention was diagnostic.
Main Outcome Measure:
Clinically reliable pure-tone audiograms were obtained in 101 patients (50.9%) for comparisons of electrocochleography and behavioral thresholds.
Results:
The mean differences between electrocochleography and behavioral thresholds at 0.5, 1, 2, and 4 kHz were less than 6 dB. Ninety-seven percent of the results were within +30 to -30 dB, and approximately 86% of the results were within +20 to -20 dB. Good correlation coefficients of 0.83, 0.84, 0.91 and 0.88 were found between electrocochleography and behavioral thresholds at 0.5, 1, 2, and 4 kHz, respectively. Only 2 of 395 ears were complicated postoperatively by suppurative otitis media. There were no anesthetic complications.
Conclusion:
Round window electrocochleography can be used in conjunction with other audiometric methods for threshold estimations in the difficult-to-test children and as part of the preoperative assessment for cochlear implantation.