Related Experiment Video
Updated: Aug 15, 2026

Radiolabeling and Quantification of Cellular Levels of Phosphoinositides by High Performance Liquid Chromatography-coupled Flow Scintillation
Published on: January 6, 2016
Flow-injection chemiluminescent determination of glycerol-3-phosphate and glycerophosphorylcholine using immobilized
M Yaqoob1, A Nabi, M Masoom-Yasinzai
1Department of Chemistry, University of Balochistan, Quetta-Pakistan.
Abstract:
Flow injection procedures with immobilized enzyme mini-columns are described for the determination of glycerol-3-phosphate, and glycerophosphorylcholine with chemiluminescent detection. The hydrogen peroxide produced on-line is coupled with a luminol (5-amino-2,3,-dihydro-1,4-phthalazinedione) peroxidation chemiluminescent system. The detection limits for glycerol-3-phosphate and glycerophosphorylcholine are 5 x 10(-7) M and 1 x 10(-6) M respectively with r.s.d. < 2%. The sample throughput is 40/h. The immobilized enzyme columns did not show any deterioration in activity after usage for 3 months.