Reliability of the Medtrac MDI Chronolog
J W Brueckner1, P Marshik, J Sherman
1College of Pharmacy, Department of Pediatrics, University of Florida, Gainesville 32610-0486, USA.
Background:
The Medtrac MDI Chronolog is an electronic device for monitoring adherence to metered-dose inhalers. It replaces previous models of the Nebulizer Chronolog and uses a different mechanism of recording actuations.
Objective:
This study was carried out to determine whether the new model can accurately record and report the date, time, and number of metered-dose inhaler actuations.
Methods:
Four canisters of beclomethasone (Beclovent) were discharged through four Chronologs with fresh batteries at a rate of 1, 2, 4, or 8 times twice daily for 7 days. Four additional canisters were used as controls and discharged simultaneously through the standard actuator. The weight of all canisters and Chronolog battery voltage were measured before and at the end of the 7-day experiment. The data retrieved from the Chronologs were compared with the information recorded manually during each discharge.
Results:
The loss in canister weight was consistent for the number of puffs discharged from all four Chronolog units and controls. However, the accuracy of the Chronologs in recording the number of actuations varied between 50% and 100%. The largest amount of data loss occurred with the unit used to discharge 8 puffs and was associated with a dead battery at the end of the 7-day trial. For actuations that were retrievable, the Chronologs accurately recorded the date and time.
Conclusions:
Unexpected battery voltage drain and other mechanical problems that we encountered may cause data loss and the false appearance of missed doses. Thus the units that we tested were not sufficiently reliable to monitor patient adherence.
More Related Videos
14:52Recording Brain Electromagnetic Activity During the Administration of the Gaseous Anesthetic Agents Xenon and Nitrous Oxide in Healthy Volunteers
Published on: January 13, 2018
06:22Surfactant Depletion Combined with Injurious Ventilation Results in a Reproducible Model of the Acute Respiratory Distress Syndrome (ARDS)
Published on: April 7, 2021
Related Concept Videos
Guidelines for Nursing Documentation II
Timely documentation is crucial to ensure continuity of care for patients. Any delays in recording or reporting medical information can result in medical errors and even adverse patient outcomes. From medication administration to diagnostic test results, every detail must be accurately and promptly documented to provide the best possible care for patients.
Methods of Documentation VI: Case Management Model
For example, a patient with a chronic illness...
Acute Respiratory Failure-I
Definition: It is defined by specific criteria based on blood gas measurements. Hypoxemia happens when the partial pressure of oxygen (PaO2) falls below 60 mmHg. At the same time,...
Acute Respiratory Failure-II
The underlying physiological abnormalities that contribute to hypoxemic respiratory failure include:
Acute Respiratory Failure-III
Acute Respiratory Failure-V
Ensure that patients are monitored continuously for their response to therapy, including changes in...